| Intended Use |
For self monitoring of blood glucose |
| Method |
Amperometric glucose biosensor |
| Physical Characteristic |
35-38mm(L) × 6mm(W) × 0.8mm(H) |
| Material |
Base Material :PET |
| |
Electrode : Carbon |
| Chemical Components |
Enzyme : Glucose oxidase |
| |
Mediator : Potassium ferricyanide |
| |
Others : Non-reactive ingredient |
| Function |
Sample Type : Capillary whole blood finger tip or AST |
| |
Sample Volume : 0.7 uL |
| |
Sampling Method : Capillary |
| |
Reaction Time : 7 seconds |
| Performance |
Linear Range : 20~600 mg/dL (1.1~33.3 mmole) |
| |
Precision : <5 % with respect to control solution |
| |
Accuracy : >95% fall in Zone A and B (Blood EGA test) |
| Stability |
Operation : 10'C ~ 40'C , RH<85% |
| |
Storage : 4'C ~ 40'C, RH<85% |
| Validity |
Can Pack : 90 days from can opened , or 18 months from production date |
| |
Individual Foil Pack : 18 months from production date |
| Warning : |
Don't use expired strip |
This strip is only used with Advocate Meters |
Close the can immediately every time when a strip is taken out |
Fluoride: Cannot put fluoride into blood as the preservative |
Iodoacetic acid: It is impossible to put Iodoacetic acid into blood as the
preservative as well |
New born baby: New born baby are not recommended to do this test |
Reductant: High density of Vitamin C or uric acid may influence
glucose value |
Hematocrit: Hematocrit above than 60% may cause lower glucose
value |
Hematocrit: Hematocrit lower than 20% may cause higher glucose
value |